EcoPhysics nCLD822M

The nCLD 822 M analyzer is the next generation in two-channel high precision nitrogen oxide measurement.EcoPhysics nCLD822M

Range: 5 ppm – 5’000 ppm
Minimum detectable concentration : 0.12 ppm
Parameters: NO, NO2 , NOx

Modular Gas Analyzer
The nCLD 822 M analyzer is the next generation in two-channel high precision nitrogen oxide measurement. Unique in speed and reliability, the nCLD 822 M is modular designed and capable of simultaneously measuring NO, NO2 and NOX. The analyzer’s expandable capabilities allow assessment of hot and humid gas sources without additional cooler. The new and intuitive graphical user interface also individually displays and connects to other instruments’ data.

nCLD – A New Generation
The nCLD 822 M includes everything for measurement of NO, NO2 and NOX. The fully revised detector-block, the enhanced gas flow paths and the improved pressure as well as temperature independence of the nCLD 800 Series instruments allow for even lower detection limits. Overall stability and reliability are lifted to a new level. The optional electro-mechanical bypass system balances out even fastest pressure variations occurring in the sample flow. Furthermore, the analyzer is adaptable to numerous non-standardized applications. The calibration of the unit runs quickly and automatically, with all necessary data available anywhere and at any time.

User Friendliness
The new touch sensitive graphical user interface enables the user to individually adjust the instrument operation and data management according to his/ her needs and applications. Multiple digital in- and outputs guarantee a maximal connectivity for your remote operation, control and maintenance of the nCLD 822 M, ensuring unsurpassed precision and reliability while all necessary data is readily available anywhere and at any time.

Compact, Modular and Intelligent!
The nCLD 822 M is manufactured in a new compact and modular layout, in which each essential component of the chemiluminescence analyzer hosts its own CPU and interacts with other CPUs by BUS-communication. This assembly increases accessibility and serviceability by reducing wiring and piping. The measurement principle will conform to the standard method for NOX-detection in stationary source emissions (EN 15267).

Analyzer type

dual chamber CLD with cooled PMT for measurement of NO, NO2 and NOX

Measuring ranges

four freely selectable ranges from 5 ppm – 5000 ppm

Min. detectable concentration*

0.12 ppm

Noise at zero point (1σ)*

0.06 ppm

Lag time

<3 sec

Rise time (0-90%)

<1 sec

Temperature range

5-40 °C

Humidity tolerance

5-95% rel. h (non-condensing, ambient air and sample gas)

Dry air use for O3 generator

internally generated (no external supply gas required)

Sample flow rate

1.0 l/min

Input pressure

600 – 1’200 mbar abs.

Power required

350 VA (incl. membrane pump and scrubber)

Supply voltage

100-240 V/50-60 Hz

Interface

USB(3x), HDMI, Bluetooth, RS232 (w/o 9pin connector), LAN, WLAN

Dimensions

height: 133 mm (5¼”)

width: 450 mm (19”)

with molding: 495mm

depth: 540 mm (21.2”)

Weight

23 kg (51 lb)

Delivery includes

nCLD 822 M analyzer, power cable, USB-LAN adapter

Standard

nCLD 822 M: NO, NO2, NOX analyzer

· M – metal converter

Options

· V1 – single calibration valve

· V2 – two calibration valves for pressurized calibration (zero & span / 2-3 bar)

· – hot tubing

· – electro-mechanical pressure regulation

· USB-RS232 9pin connector

· analog output (external box): 0-10 V ; 4-20 mA into 500 Ω max.

 

 

*Depending on filter setting.

Connectivity properties are country-specific.

ECO PHYSICS reserves the right to change these specifications without notice.